MEETING WITH TIS-HIGH VOLTAGE (12-12-03)
Present: G. Salomon, J-F. Kepinski, I. Podadera.
Location: building 275.
Subject: Installation of ion source and RFQCB high
voltage at building 275.
- A message has to be put in the crane control advertising
people not to use the crane over the HV cage and the experiment.
- All the experimental area will be surrounded for HV
planes. It is important that these panels are very rigid (the dimension of the
grid is not so important…).
- One of the door should be placed in front of the door of
the HV cage for the Ion source power supplies.
- For the work to put the panels is possible to contact
77777 (electrical works).
- The standard distance between 2 HV devices in 60 kV in
air is 180 mm. The distance is 600 mm to a vertical metal grid.
- In the enter of the building should already appear the
warning of HV danger.
- All the equipment has to be connected to the building
ground.